Monitoring of Synthesized Compound with ELSD

 

The following data  shows usefulness of the SofTA ELSD compared with UV detector. As G-OMe does not have UV absorbance, it is difficult to monitor this synthesis reaction with UV detector.

G-OMe(Glycine methylester) + Boc-F (Butoxy phenylalanine) = Boc-F-G-OMe

Column; BETASILC18 4.6 x 150mm L

Col. Temp.; Ambient

Mobile Phase; A_H2O, B_MeOH

Gradient; 0 min 40 % B, 7 min 100 % B, 15 min 100 % B

Flow Rate; 1 ml/min

SofTA conditions; Blue line

SC temp.; 45 C

DT temp,; 60 C

FLT; 0

UV; 254 nm (0.001 au 10mV) Red line

Copyright © 2008 SofTA Corporation.  All rights reserved.
Revised: 05/01/08